Technical Information:
| Crystal EX Application | Trace | HPLC | Bio |
| Reagent Preparation | ✓ | ✓ | ✓ |
| Flame AAS | ✓ | ✓ | ✓ |
| Autoclaves | ✓ | ✓ | ✓ |
| ICP-MS | ✓ | ✓ | ✓ |
| ICP-OES | ✓ | ✓ | ✓ |
| İon Chromatography | ✓ | ✓ | ✓ |
| Liquid Chromatography (HPLC) | ✓ | ✓ | |
| Gas Chromatography (GC) | ✓ | ✓ | |
| TOC Measurements | ✓ | ✓ | |
| Flow Cytometr0079 | ✓ | ||
| Cell and Tissue Culture | ✓ | ||
| Molecular Biology | ✓ |
Feed Water Requirements for Pure Water System;
· Feed Water: Potable Tap Water
· Pressure: 0,4–4,0 bar
· Conductivity: <1500 μS/cm
· Temperature: 5–35 °C
If the feed water does not meet specifications, additional pre-treatment solutions are available.
| Crystal EX Technical Specifications | Trace | HPLC | Bio |
| Grade 1 Water Resistivity at 25°C | 18.2 MΩ x cm | 18.2 MΩ x cm | 18.2 MΩ x cm |
| Grade 1 Water Conductivity at25°C | 0.055 μS/cm | 0.055 μS/cm | 0.055 μS/cm |
| Grade 2 Water Resistivity at 25°C | >10 MΩ x cm | >10 MΩ x cm | >10 MΩ x cm |
| Grade 2 Water Conductivity at 25°C | < 0.1 μS/cm | < 0.1 μS/cm | < 0.1 μS/cm |
| TOC Level | < 10 ppb | < 5 ppb* | < 5 ppb* |
| Rnase | - | - | < 0.01 ng/mL |
| Dnase | - | - | < 4 pg/mL |
| Bacteria | < 0.01 CFU/mL | < 0.01 CFU/mL | < 0.01 CFU/mL |
| Endotoxins | < 0.15 EU/mL | < 0.15 EU/mL | < 0.001 EU/mL |
| Particles >0.22 μm | < 1/mL | < 1/mL | < 1/mL |
| Nominal flow, Pure Water (to storage tank) | 10 L/h | 10 L/h | 10 L/h |
| Dimensions (WxDxH), cm | 40*35*55 cm | 40*35*55 cm | 40*35*55 cm |
| System Weight, kg | 17 | 18 | 19 |
| Operating Weight, kg | 24 | 25 | 26 |
| * In appropriate operating conditions <2 ppb, otherwise normally <5 ppb. | |||
Consumables Required for Crystal EX;
| Part No. | Description | Explanation |
| 10320 | Replacement pre-filter, Crystal EX | |
| 10310 | Replacement deionization module | |
| 10030 | Replacement polishing module | |
| 10017 | Sterilization UV bulb | Bio configuration |
| 10018 | Replacement photooxidation UV bulb | HPLC and Bio configuration |
| 10013 | Replacement 0.22 µm dispense microfilter, sterile | Trace and HPLC configuration |
| 10120 | Replacement ultrafilter | Bio configuration |